Especificações
2. Scope of application: adult, child
3. Gas source:
O2: 0.27~0.55MPa
N2O: 0.27~0.55MPa
4. Flow meter
O2 : 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
Air : 0.1~10L/min
5. O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
6. Flow rate of rapid oxygen supply:25~75L/min
7. Low oxygen pressure alarm:There will be sound alarm when the oxygen pressure < 0.2MPa
8. Vaporizer:it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,
Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
9. Respiratory circuit:working mode: all-close, semi-close, semi-open APL: 0.5~7 kPa
10. Respiratory Bellows:bellows for adults, tidal volume range:0~1500 ml
bellows for children, tidal volume range:0~300 ml
anyone can be chosen for application as required by the customer
Ventilator:
1. Display mode:High-definition 8.4” TFT color LCD screen display
2. Ventilation mode:IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH(IMV, VCV, PCV)
3. Ventilation parameters:
Tidal volume:50~1500 ml (20~1500 ml)
Rate:1~99 bpm
SIMV rate:2~20 bpm I:E 2:1~1:8(4:1~1:8)
Inspiratory trigger pressure:-1.0~2.0 kPa PEEP 0~2.0 kPa
Pressure range:1.0~6.0 kPa
SIGH:1.5 times the inspiratory time (70-150 adjustable)
Inspiratory platform:0~50%
4. Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate,
total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform
(autonomous inspiratory rate, lung compliance, airway resistance,flow-volume loop, pressure-volume loop)
5. Monitoring of oxygen concentration:21%~100%
6. Security Alarm System
Oxygen Concentration alarm upper limit setting range 21%~100%
low limit setting range 10%~80% Airway pressure alarm upper limit setting range 0.3~6.0 kPa
low limit setting range 0.2~5.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min