- Description
Detector:
· Number of detector rows: 8
· Number of detectors per row: 704
Tube:
· Anode heat storage capacity: 1.5MHU
· Anode heat dissipation:441KHU/Min
· Small focal spot size: 0.7mm×0.8mm
· Large focal spot size: 1.2mm×1.4mm
Specification of CT Scan Machine
1 | Gantry |
Aperture:70cm Manual control from POD at console Gantry cooling: air Gantry rotation: belt-driven Slip ring type: low voltage Positioning system: 3D lasers |
|
2 | HV Generator |
Power rating: 32kW MA range: 10-300mA KV range: 70kV, 80kV, 100kV, 110kV, 120kV, 140kV. |
|
3 | Tube |
Anode heat storage capacity: 1.5MHU Anode heat dissipation:441KHU/Min Small focal spot size: 0.7mm×0.8mm Large focal spot size: 1.2mm×1.4mm |
|
4 | Detector |
Number of detector rows: 8 Number of detectors per row: 704 |
|
5 | Table |
Maximum motion range:1600mm Max. speed in horizontal direction:150mm/s Motion tolerance in horizontal direction (200 kg):±1mm Table elevation range:440mm Max weight load:450lb(205kg) Foot pedal control, table horizontal motion |
|
6 | Console |
Uninterrupted power supply UPS (3KVA) to prevent sudden power outages and ensure the system stability Memory:8G Hard disk:850 GB Frequency:2.0 GHz Liquid crystal display of high resolution, displaying matrix (M * M): 1920×1200 Progressive scan display DICOM 3.0 network interface Burn mode:DVD DICOM 3.0 print interface DICOM 3.0 output and input interface Auto-film Auto-voice and integrated intercom for communication between operator and patient Include DICOM PRINT,DICOM STORE,DIOCM QUERY,DICOM RETRIVE,WORKLIST Include HIS & RIS interface Raw data can be reconstructed on the workstation |
|
7 |
Scan and Image Reconstruction |
The fast rotation time: 0.75s/360° Longest scanning time:100s Longest scanning range:1200mm Scout scan direction: anteroposterior, lateral Pitch range: 0.25-1.75 Maximum number of slices per rotation:16 slices Thinnest image slice thickness in 16-slice mode:1.25mm Thinnest reconstructed slice thickness in 16-slice mode:1.25mm Nominal slice thickness for axial scan: 1.25 mm,2.5 mm, 5 mm,10 mm Nominal slice thickness for helical scan: 1.25 mm,2.5 mm,3.75 mm,5.0 mm,7.5 mm,10 mm Parallel image processing Simultaneous Reconstruction: Can reconstruct and reorganize the images simultaneously with a variety of options. Different reconstruction options are available within the scan protocol. Scanning FOV: 43cm Image reconstruction matrix:1024×1024 Image display matrix:1024×1024 |
|
8 | PDU |
Input voltage: 3-phase 380V, 50Hz Power: 50kVA |
|
1 | System Software |
Beam hardening artifact correction software Posterior cranial fossa optimization Motion correction reduces motion artifact Pediatric protocol NDI (NanoDose Iteration) Filming print software Remote maintenance system |
|
10 | Accessory KIT |
Patient Table Pad, Head Holder, Head Holder Cushion, Phantom. | |
11 | Clinical Application |
Include Volume Reconstruction (VR), Multi-planar Reconstruction(MPR), Curved Planar Reconstruction (CPR), Surface Shaded Display(SSD), MIP,MinP. | |
12 | Option |
UPS,Printer,High Pressure Injector,FOV 500mm,Workstation,Abdominal fat analysis,The orthopaedic module |