- Description
Specifications
* Single diagnostic table and single X-ray tube with rotated anode.
* X-ray tube capacity protection, anode starting-up protection and radiograph time-limited protection devices are provided.
* The surface of the bed can move lengthways and crosswise, electromagnetism locking.
* The X-ray tube assembly can rotate the cross arm, and be localization anywhere and vertical radiograph frame is provided.
Technology Parameters
Description | Technology Parameters | |
Power Supply | Voltage | A.C.380V / 220V (±10%) |
Frequency | 50Hz ± 1Hz | |
Capacity | 20KVA | |
Power Supply Resistance | A.C.380V≤1.0Ω; A.C.220V≤0.3Ω | |
Maximum tube voltage | 120KV; 125KV | |
Maximum tube currency | 200mA | |
Radiograph | Tube voltage | 40~120KV; 40~125KV |
Tube currency | Small Focus: 50mA; 100mA; Big Focus: 100; 150; 200mA | |
Time | Divided into 23 grades from 0.04~6.3s | |
High Voltage Generator | Capacity | 20KVA |
X-ray Tube | Model | XD 51—20、40 / 125 |
Focus | Small Focus 1×1mm2 Big Focus 2×2mm2 | |
Rotated Speed of anode | 2800r / min | |
Table for radiograph | Length×width×height | 2000×760×680(mm) |
Movable scope of the bed | Lengthways: ±450mm; Crossways: ±110 | |
Pillar moves lengthways | 1000mm | |
Distance from surface to film | ≤60mm | |
Bulb tube moves up and down | 1100mm~1760mm | |
Radiography item | 10〞×12〞 11〞×14〞 12〞×15〞 | |
Beam Limited Device | Manually single automatic multi-optional |