- Description
Specifications
Automatic self-testing
Biphasic energy output
CPR metronome
CPR Coaching (Optional)
Adult and Pediatric defibrillation
Extensive voice and visual prompts foe the operator
Audible Prompts 19 audible prompts
Visual Prompts LED
Controls six buttons - On/Off, Shock, and 4 software-configurable buttons
Output Energy Accuracy ±15% into any impedance from 25 to 175Ω
The Maximum Voltage 1800±20V
Output disabled when the PATIENT impedance is outside limits 20Ω to 200Ω
Operating Temperature 32°F to 122° F; 0°C to 50° C
Storage Temperature -22°F to 140° F; -30°C to 60° C
Humidity 10% to 95% relative humidity, non-condensing
Vibration MIL Std. 810F, Min Helicopter Test
Shock IEC 68-2-27; 100G
Altitude -300 to 15,000 ft.; -91m to 4573m
Particle and Water Ingress IP-34
Transient operating conditions: temperature -4°F to 122° F; -20°C to 50° C