- Description
Specifications
Display screen:
LCD with high brightness
Sample volume:
10μL~50μL
Interface:
RS-232 serial port, able to be connected to computer
Certificates:
Recognized by TUV and CMD
Light Source: 6V 10W, halogen lamp of more than 2000 hours’life time
Spectrometer: interference filter.
Wavelength: 340nm,405nm,505nm,546nm,578nm,620nm, and one wavelength free.
CML precision: ±2nm
FWHM: ≤10nm
Stray light:≤0.5%(absorbance≥2.5)
Absorbance range: -0.214~3.000
Absorbance linearity: ±1%
Absorbance repeatability: ±0.001
Absorbance resolution: 0.0001
Precision: ≤1.0%
Carry over: ≤1%
Temperature: 25°C,30°C,37°C,±0.3°C
Flow cell:permanent quartz flow cell, minimal volume is 32μL, and disposable flow cell
is selectable
Sampling system: pressure auto releasing pumps ensure the accuracy of suction volume
Operational system: VISTA,Windows XP,Windows2000
Test Method: End point,Fixed time,Kinetic,Factor,Muli-standards,Bichromatic,Dual wavelengths
Parameter Setting: Patients’profile and calculated item can be set easily
Absoption volume: 200μL~3000μL
Sample volume: 10μL~50μL
Reaction curve: real time display the reaction curve
Memory: Able to store more than 200 parameters while 55 parameters of them are stablized ;
Able to save more than 15000 test results and auto save the results if the power is off during working
QC: auto draw the QC figure
Reagent: Open reagent system
Display: LCD with high brightness
Input: press button
Printer: Inner thermal printer,able to beconnected to external printer
Report content: absorbance,concentration,activity, detailed patient information and
abnormal results marking
Interface: RS-232 serial port, able to be connected to computer
Working Condition: Temperature:10°C~30°C;Relative Humidity:20~80
Voltage: AC 85-264V,auto adapt according to the input voltage
Power : ≤100W
Quality System: Recognized by TUV and CMD