Sua posição: Página inicial > Ventilator & Anesthesia machine > Anesthesia machine

High-strength MY-E008 Movable Anesthesia machine

Código do item: MY-E008
    Selecione as informações desejadasX
Custo de envio:
para
Estimativa de Entrega:
Selecione o país do qual deseja enviar
+
-
Unit/Units 50 em Stock
  • Descrição
Especificações
Specification:
1. Main body :High-strength engineering plastic rack, light,beautiful and corrosion resistant
2. Scope of application:adult, child
3. Gas source:O2: 0.27~0.55MPa  N2O: 0.27~0.55MPa
4. Flow meter:O2: 0.1~10L/min    N2O: 0.1~10L/min
5. O2, N2O linkage and N2O Stopper
    When using nitrous oxide, oxygen concentration>25%;
    When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.
 
6. Flow rate of rapid oxygen supply :25~75L/min
7. Low oxygen pressure alarm :There will be sound alarm when the oxygen pressure < 0.2MPa
 
8. Vaporizer:it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.
 
9. Respiratory circuit :working mode: all-close, semi-close, semi-open
                                 releasing pressure: 0.5~7 kPa
10. Respiratory bellows:bellows for adults, bellows for children
      tidal volume range:0~1,500 ml
 
Ventilator:
1. Display mode : High-definition 5.7” LCD screen display
2. Ventilation mode:IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH
3. Ventilation parameters:
    Tidal volume:  50~1500ml
    Rate:2~99bmp
    SIMV rate: 2~20bpm
    I:E :2:1~1:8
    Inspiratory trigger pressure: -1.0~2.0 kPa
    PEEP:0~2.0 kPa
    Pressure range:1.0~6.0 kPa
    SIGH:1.5 times the inspiratory time
4. Parameters for ventilation monitoring  tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure.
5. Monitoring of oxygen concentration :(21%~100%)
6. Security Alarm System
     Oxygen Concentration alarm :upper limit setting range (50%~100%)
     low limit setting range :(15%~50%)
     Airway pressure alarm:upper limit setting range  1.0~6.0 kPa
     low limit setting range:0.4~2.0 kPa
     Per-minute ventilation volume alarm  upper limit setting range:  3.0~30L/min
     low limit setting range:1.0~10L/min
     Sustained high-pressure alarm: it will give alarm when stress have consistently
     been higher than 2.5 kPa
     Suffocation alarm:it will give sound and light alarm if there is
     no tidal volume input for 15 seconds Power alarm
7. Power                      
 AC 220V±10%     50Hz±1Hz   (UPS, Storage battery)