- Description
Specifications
1. Self-checking function;
2. Carbonate dialysis;
3. Double needle dialysis;
4. Liquid level detector;
5. Bubble detector;
6. Blood leakage detector;
7. Temperature and electric conductivity monitoring;
8. Arterial pressure, venous pressure and transmembrane pressure monitoring;
9. Rolling blood pump;
10. Heparin pump;
11. Dehydration amount controlled by capacity;
12. Automatic disinfection cleaning program;
13. Stand-by power of blood pump in case of power failure;
14. Information display function of display screen.
Technical Data
General data:
Dimensions 1380 x 380 x 400 mm (H x W x D);
Weight approx. 88 KG;
Water supply:
Water inlet pressure 1.5-6.0 bar
Water inlet temperature 5℃~32℃
Max. drain height 1m
Concentrate supply
Supply pressure 0 to 2 m suction height
Electrical data
Power supply AC220V±10%, 47-63HZ
Current consumption max. max. 10 A
Extracorporeal circuit
Arterial pressure monitoring
Display range -380 mmHg to +400 mmHg
Accuracy ±10 mmHg
Resolution 1 mmHg
Venous pressure monitoring
Display range -180 mmHg to +600 mmHg
Accuracy ±10 mmHg
Resolution 1 mmHg
Transmembrane pressure monitoring
Display range -180 mmHg to +600 mmHg
Accuracy ±20 mmHg
Resolution 1 mmHg
Arterial blood pump
Blood flow range 30 to 600ml/min in 8 mm bloodline systems
Accuracy ±10%
Air bubble pump
By ultrasound transmission, additional optical monitoring in venous clamp.
Heparin pump:
Delivery range 0ml/h to 10ml/h
Bolus function max. 20 mL per bolus
Syringe size 10, 20, 30, 50ml
Dialysis fluid circuit
Dialysis fluid flow range
Selectable 100-800 mL/m