- Description
Specifications
50ml syringe:0.1~1500ml/h;
30ml syringe:0.1~900.0ml/h;
20ml syringe:0.1~600.0ml/h;
10ml syringe:0.1~300.0ml/h
The range of limiting volume
0~999.9ml
The range of injected volume
0~9999.9ml
Accuracy of flow rate
Within ±3%(Choose the syringe of specified manufacturer correctly. This accuracy of flow rate is not including the errors of syringe itself.)
Purge rate
Purge rate of 50ml syringe:1500ml/h;
Purge rate of 30ml syringe:900.0ml/h;
Purge rate of 20ml syringe:600.0ml/h;
Purge rate of 10ml syringe:300.0ml/h
Alarm and warms
The pump can give out alarming when injection soon finish, injection finishing, injection of limited amount finished, injection is blocked, syringe falls off, syringe is improperly installed, setting error, AC power has been pulled out, low battery and so on.
Injection blocked
Choose in the three kinds of occlusion alarm pressure according to the clinical requirements(High,Medium,Low)
KVO rate
1ml/h
Fuse specification
F1AL/250V ,two fuses(installed inside pump)
Operating power
AC100~240.0V, 50/60Hz, Power:≤30VA
Internal rechargeable battery: DC 11.1V,capacity≥1600mAh. The device could work constantly more than 4 hour at medium speed running mode, after the battery charged for 10 hours.
Operating environment
temperature:+5~+40°C, relative humidity:20~90%
Storing environment
temperature:-30°C~+55, relative humidity:≤95%, ( No condensation state )
Atmospheric pressure
860hPa~1060hPa
Safety level
Class II with insulated enclosure, internal power supply, BF-type continuous normal running equipment.
IP grade
IPX1
Dimensions
280(L)×210(W)×130(H)(mm)
Weight
2.2kg
Accessories
Power line,operating manual,warranty card,product certification.